http://www.anritsu.com 131
DIGITAL LINK MEASURING INSTRUMENTS
ERROR DETECTOR
MP1776A
100 MHz to 12.5 GHz
2
Supports Measurement for up to 50 Gbit/s System (Installed with 4 Channels)
MP1776A is an error detector housing four error detectors that can
measure error up to 12.5 Gbit/s. It has four-channels independent
measurement mode, two-channels or four-channels combined mea-
surement mode and be used for development, manufacturing and
maintenance of transmission systems and modules from 12.5 Gbit/s
to maximum 50 Gbit/s.
Features
• Max. 4-channels in one box
• Independent measurement of 4-channels
• PRBS patterns from 2
7
– 1 to 2
31
– 1
• Max. 32 Mbit programmable pattern at 4-channels combined mode
(corresponding six frames of STM-256/ STS-768)
• Burst data BER measurement for optical circulating loop test
• Good operability by GUI
• Display 4-channels measurement results on screen
Specifications
• MP1776A (main frame), MU177601B (12.5 Gbit/s Error Detector)
Operating frequency 100 MHz to 12.5 GHz
PRBS pattern: 2
n
– 1 (n: 7, 9, 11, 15, 20, 23, 31)
Zero substitution pattern: 2
n
(n: 7, 9, 11, 15), consecutive zero-pattern can be inserted up to pattern length – 1
Programmable data
Measurement pattern Independent: 2 to 8,388,608 bits
2-channels combined: 4 to 16,777,216 bits
4-channels combined: 8 to 33,554,432 bits
Logic inversion: Positive/negative switching possible
Measurement mode Independent
∗
1
, 2-channels combined
∗
2
, 4-channels combined
∗
3
Synchronization method Normal, frame
Error detection mode Insertion, omission, total
Error ratio: 0.0000 x 10
–16
to 1.0000 x 10
0
Error count: 0 to 9,999,999, 1.0000 x 10
7
to 9.9999 x 10
16
Measurement items Clock frequency:
0.1 to 12.5 GHz (independent), 0.2 to 25 GHz (2-channels combined), 0.4 to 50 GHz (4-channels combined)
∗Resolution: 1 kHz, accuracy: 10 ppm ±1 kHz
Sync threshold value Internal, 10
–n
(n: 2, 3, 4, 5, 6, 7, 8)
Auto search function Supported
Input waveform: NRZ
Input amplitude: 0.5 to 2.0 Vp-p
Threshold voltage: –3.000 to +1.750 V (1 mV step)
Data input Termination condition: GND/–2.0 V
Input impedance: 50 Ω
Connector: APC-3.5
Number of input: 1 (MU177601B 12.5 Gbit/s Error Detector Unit)
Input level: 0.5 to 2.0 Vp-p
Input waveform: Square wave only (<0.5 GHz, Duty: 50%), Sine or square wave (≥0.5 GHz, duty: 50%)
Clock delay: ±500 ps (1 ps step)
Clock input Polarity inversion: POS/NEG inversion selectable
Input impedance: 50 Ω
Connector: APC-3.5
Number of input: 1 (MU177601B 12.5 Gbit/s Error Detector Unit, up to 4 channel can be added.)
Continued on next page