Anritsu ML2430A Manual de usuario Pagina 217

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 271
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 216
Status
Register
Control
This function demonstrates how to use the Status Registers to provide synchroni
-
zation.
Uses the TR2 (trigger with settling) command to make a reading.
Function GetTR2Reading (channel) As Single
make space for the result
Dim result As String
result = String$(10, 0)
Send Status Register setup command + TR0 hold trigger mode
Call DLLsend(0, 13, “*SRE 16; TR2 1", 14, NLend, ibsta%, iberr%, ibcntl&)
Set loop flag
Value = -256
Do
Loop until SRQ is asserted.
Do
Call DLLTestSRQ(0, SRQ%, ibsta%, iberr%, ibcntl&)
Loop Until SRQ%=0
SRQ asserted, read the ML2430As status register
Call DLLReadStatusByte(0, 13, status_byte%, ibsta%,
iberr%, ibcntl&)
Check if it is the ML2430A which is requesting
service (SRQ bit + MAV bit)
If (status_byte% And 80) = 80 Then
'
It is the ML2430A, read back value
Call DLLreceive(0, 13, result, 10, STOPend,
ibsta%, iberr%, ibcntl&)
Value = Val(result)
End If
Loop Until Value <> -256
GetTR2Reading = Value
End Function
6-130 ML2430A OM
PROGRAMMING EXAMPLES GPIB OPERATION
Vista de pagina 216
1 2 ... 212 213 214 215 216 217 218 219 220 221 222 ... 270 271

Comentarios a estos manuales