I Trigger without settling.
Remarks:
Triggers a single reading which is added to the internal digital filter and the
updated filter power level is returned on the GPIB. After an I command, the
instrument returns to standby mode (H).
OI Identification
Remarks:
Ask for identification of current operating mode. Responds with “HP436.
R Free run mode
Remarks:
Sets the ML2430A Series back into free run mode to continuously take
measurements and output data.
T Trigger with settling
Remarks:
Triggers a new series of readings; enough to update the internal digital fil-
ter for a noise free reading at the current power level. The value is then re-
turned on the GPIB and returns to standby mode (H).
V Free run mode with settling
Remarks:
Sets the ML2430A Series back into free run mode to continuously take
measurements and output data after running a settling routine.
Z Zero sensor
Remarks:
Zero out the noise from the sensor. When zeroing the ML2430 in HP 436
emulation mode, the 'Z1T' sequence followed by the '9+DI' described in
the HP 436 manual must be followed.