Remarks:
Triggers a new series of readings; enough to update the internal digital fil
-
ter for a noise free reading at the current power level. The value is then re
-
turned on the GPIB and returns to TR0 standby mode.
Related
Commands:
TR0, TR1, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
TR3 Trigger free run
Syntax:
TR3
Remarks:
Sets the ML2430A Series back into free run mode on both channels.
Related
Commands:
TR0, TR1, TR2, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
ZE Zero sensors
Syntax:
ZE
Remarks:
Zero all connected sensors. The ML2430, when emulating the HP 437,
may take longer to Zero a sensor than the HP 437 itself. When performing
a zero the status byte should be used to identify when ZEROing is com-
plete.
6-116 ML2430A OM