When selecting GPIB emulation modes, the instrument configures itself to
the preset conditions of the instrument to be emulated. For example, when
selecting HP 438A emulation, the front panel menus pass through the pre
-
sets for the HP 437B (which presets sensor A to dBm) then selects HP
438A emulation (which presets sensor A to Watts).
NOTE
This command must be entered using the 488.2 format;
that is, EMUL<ws><mode> (<ws> = white space).
MCA Set the cal factor value at the specified memory location in
dBm.
Syntax:
MCA<mem><;><value>
mem:
Memory location 1 to 30.
value:
Cal factor value in dBm.
Remarks:
Set the cal factor value at memory store address <mem> to <value> dBm.
MCC Clears the cal factor value at the specified memory loca-
tion.
Syntax:
MCC<mem>
mem:
Memory location 1 to 30.
Remarks:
Clears the cal factor value at memory store <mem> to 0.0 dBm.
MCO Clears the offset value at the specified memory location.
Syntax:
MCO<mem>
mem:
Memory location 1 to 30.
Remarks:
Clears the offset value at memory store <mem> to 0.0 dBm.
MCQ Clears the frequency value at the specified memory loca
-
tion.
Syntax:
MCQ<mem>
mem:
Memory location 1 to 30.
Remarks:
Clears the frequency value at memory store <mem> to 0.1MHz.
ML2430A OM 6-87
GPIB OPERATION ML4803A EMULATION COMMANDS