SECTION 6 GPIB
6-34
:AMPLitude[ :OUT]:RELative <Boolean>
Function Turns ON or OFF the relative level display mode.
Parameter <Boolean> = ON or 1
OFF or 0
Unit <Non term>
Restriction None
:AMPLitude[ :OUT]:RELative?
Function Reads out the state of relative level display mode.
Response Relative level display mode = OFF
ON
Restriction None
:AMPLitude[ :OUT]:OFFSet <Numeric> <Ampl term>
Function Sets the output level offset value.
Parameter <Numeric> = 55 dB to 55 dB
Unit <Ampl term>
Restriction Unit is dB, only.
:AMPLitude[ :OUT]:OFFSet?
Function Reads out the output level offset value.
Response Level offset value
Restriction None
:AMPLitude[ :OUT]:ISOLation <Boolean>
Function Turnd ON or OFF the isolation mode.
Parameter <Boolean> = ON or 1
OFF or 0
Unit <Non term>
Restriction None
:AMPLitude[ :OUT]:ISOLation?
Function Read out the state of isolation mode.
Response Isolation mode = OFF
ON
Restriction None
:AMPLitude[ :OUT]:SWEep <Character>
Function Executes output level sweep.
Parameter <Character> = RUN
STOP
PAUSE
CONT
Unit <Non term>
Restriction None
:AMPLitude[ :OUT]:SWEep:STARt <Numeric> <Ampl term>
Function Sets the start level for a sweep.
Parameter <Numeric> = 143 dBm to +23 dBm
Unit <Ampl term>
Retriction None