Anritsu MG3642A Manual de usuario Pagina 103

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 212
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 102
SECTION 5 MEASUREMENT
5-11
STEP PROCEDURE
7. Turn on the output of MG3641A/MG3642A , place MG3641A/MG3642A into the relative level mode,
and set the output level resolution to 1 dB.
8. Set the frequency step size of MG3641A/MG3642A to 20 kHz.
9. Turn the rotary knob to adjust the output level so that the noise output of the receiver is set to V
N
dB
obtained in step 2 each time the Edit [^] key of MG3641A/MG3642A is pressed. The value indicated by
the output level indicator is the disturbing input level separated from the desired wave by + f × n.
10. Return the frequency and output level of MG3641A/MG3642A in the state set in step 3 (frequency and
output level indicators set to +0).
11. Turn the rotary knob to adjust the output level so that the noise output of the receiver is set to V
N
dB
obtained in step 2 each time the Edit [ ] key of MG3641A/MG3642A is pressed. The value indicated by
the output level indicator is the disturbing input level separated from the desired wave by f × n.
12. The following sensitivity suppression characteristics are obtained in steps 9 and 11:
>
80604020020406080
120
100
80
60
40
20
Disturbing wave input level (dB)
Disturbing frequency offset [kHz]
Vista de pagina 102
1 2 ... 98 99 100 101 102 103 104 105 106 107 108 ... 211 212

Comentarios a estos manuales

Sin comentarios