SECTION 6 GPIB
6-33
:AMPLitude[ :OUT]:STATe <Boolean>
Function Turns ON or OFF the RF output.
Parameter <Boolean> = ON or 1
OFF or 0
Unit <Non term>
Restriction None
:AMPLitude[ :OUT]:STATe?
Function Reads out the state of RF output.
Response RF output = OFF
ON
Restriction None
:AMPLitude[ :OUT]:CONTinous <Boolean>
Function Turns ON or OFF the level continuous mode.
Parameter <Boolean> = ON or 1
OFF or 0
Unit <Non term>
Restriction None
:AMPLitude[ :OUT]:CONTinuous?
Function Reads out the state of level continuous mode.
Response Level continuous mode = OFF
ON
Restriction None
:AMPLitude[ :OUT]:SAFety <Boolean>
Function Turns ON or OFF the level safty mode
Parameter <Boolean> = ON or 1
OFF or 0
Unit <Non term>
Restriction None
:AMPLitude[ :OUT]:SAFety?
Function Reads out the state of level safety mode.
Response Level safety mode = OFF
ON
Restriction None
:AMPLitude[ :OUT]:VOLT <Character>
Function Selects the voltage display mode.
Parameter <Character> = EMF
TERM
Unit <Non term>
Restriction None
:AMPLitude[ :OUT]:VOLT?
Function Reads out the RF voltage display mode.
Response Open voltage display = EMF
Terminated voltage display = TERM
Restriction None