NOTE
The measured value is converted to dBc/Hz us
-
ing the following formula:
()
[]
dBc Hz measureddBc
RBW
Hz
/
log
=- - 10
1
At 10 kHz RBW,
()
10 40
1
log
RBW
Hz
=
,so
dBc Hz measureddBc/ =- - 40
.
23. Press the
BACK soft key and the M3 soft
key.
24. Press the
EDIT key and enter 999.970.
Press
MHz to set the M3 marker fre
-
quency to 999.970 MHz (30 kHz lower
than the center frequency).
25. Press the
DELTA (M3-M1) soft key.
26. Press the
RUN/HOLD key. Read and record
the D 3 reading.
27. Press the
RUN/HOLD key to read and re
-
cord five values, then calculate the aver
-
age of the five recorded values.
28. Substract 40 dB from the average value
and verify that the result is £ –75
dBc/Hz. (For example: –35 dBc measured
40 dB = –75 dBc/Hz.)
7. SPECTRUM ANALYZER SECOND
HARMONIC DISTORTION TEST
The following test can be used to verify the input re-
lated spurious response of the Site Master Spectrum
Analyzer.
a. Equipment Required:
·
Anritsu MG3692A Synthesized Signal
Source with options 2A, 4 and 15A, or
equilvalent
·
10 MHz Reference Standard
·
50 MHz Low Pass Filter
·
RF coaxial cable, N male to N male
b. Procedure:
1. Connect the 10 MHz reference source to
the Anritsu MG3692A Synthesized Sig
-
nal Source.
2. Connect one end of the 50 MHz Low Pass
Filter to the output of the source and the
other end to the Site Master Spectrum
Analyzer RF Input with the RF coaxial
cable.
3. Connect the external power supply
(Anritsu part number 40-115) to the Site
Master.
4. On the Site Master, press and hold the
ESCAPE/CLEAR key, then press the
ON/OFF key to turn on the Site Master.
(This sets the instrument to the factory
preset state.)
5. Turn on the 10 MHz reference source
and the Anritsu MG3692A Synthesized
Signal Source.
6. Set the MG3692A output to 40MHz CW,
with an RF output level of –30 dBm.
NOTE
Before continuing, allow a 30-minute warm up
for the internal circuitry to stabilize.
7. On the Site Master, press the
MODE key.
Use the up/down arrow key to highlight
SPECTRUM ANALZYER and press ENTER
to select spectrum analyzer mode.
8. Press the
AMPLITUDE key and the REF
LEVEL
soft key.
9. Enter –27 and press the
ENTER key to set
the Reference Level to –27 dBm.
10. Press the
SCALE soft key and enter 7
then press ENTER.
11. Press the
FREQ/DIST key and then the
CENTER soft key.
12. Enter 40 and press the
MHz soft key to
set the center frequency to 40 MHz.
13. Press the
SPAN soft key and enter 0.2.
Press the
MHz soft key to set the span to
0.200 MHz.
14. Press the
SWEEP key.
15. Press the
RBW soft key and then the
MANUAL soft key. Use the Up/down arrow
key to select
10 kHz. Press ENTER to set
the resolution bandwidth to 10 kHz and
select
BACK.
16. Press the
VBW soft key and then the
MANUAL soft key. Use the Up/down arrow
key to select
3 kHz. Press ENTER to set
the video bandwidth to 3 kHz.
17. Press the
MARKER key, then the M1 soft
key.
18. Select the
EDIT soft key, then enter 40.
Press the
MHz soft key to set M1 to 40
MHz .
4 Site Master S114C/S332C MM