a. Equipment Required:
·
InstaCal Module, part number ICN50
·
20 dB offset, Anritsu SC5270
·
6 dB offset, Anritsu SC5237
b. Procedure
NOTE
If performing full verification over the entire
frequency range with both an S114C and
S332C Site Master, perform this procedure
first on the S114C and then on the S332C.
1. Press and hold the
ESCAPE/CLEAR key,
then press the
ON/OFF key to turn on the
Site Master. (This sets the instrument to
the factory preset state.)
NOTE
Before continuing, allow a five minute warm
up for the internal circuitry to stabilize.
2. Press the
MODE soft key.
3. Use the Up/Down arrow key to highlight
RETURN LOSS, then press ENTER.
4. Press the
START CAL key. The message
“CONNECT OPEN or InstaCal TO RF
Out PORT” will appear in the display.
5. Connect the InstaCal module to the RF
Out port and press the
ENTER key.
NOTE
If this particular InstaCal module has been
used to calibrate this Site Master before, the
Site Master senses the familiar InstaCal
module and automatically calibrates the unit
using the OSL procedure.
If the Site Master senses that the
characterization data for the InstaCal module
connected to this Site Master is different than
the one currently stored, it will display soft
key options to keep or replace the InstaCal
characterization data.
Selecting the
YES soft key transfers all of the
characterization data from this InstaCal
module to the Site Master. The transfer may
take up to three minutes. This option is
preferred if this InstaCal module is to stay
with this particular Site Master. Once
completed, the data will not need to be
transfered again for this combination of Site
Master and InstaCal module.
Selecting the
NO soft key will temporarily
transfer only the portion of the characteriza
-
tion data necessary for this particular calibra
-
tion. This transfer takes aproximately 30 to 60
seconds, and will have to be repeated every
time a calibration is done using this combina
-
tion of Site Master and InstaCal module.
6. Verify that the calibration has been prop
-
erly performed by checking that the
CAL
ON!
message is displayed in the upper
left corner of the display.
7. Remove the InstaCal module from the
RF Out port and connect the 20 dB Off
-
set to the RF Out port.
8. Measure the return loss of the 20 dB Off
-
set. The level should be 20 dB, ±2dB
across the calibrated frequency range.
9. Remove the 20 dB Offset from the RF
Out port and connect the 6 dB Offset to
the RF Out port.
10. Measure the return loss of the 6 dB Off
-
set. The level should be 6 dB, ±1.2 dB
across the calibrated frequency range.
Site Master S114C/S332C MM 15