hering to connector interfaces. When not in use, keep the connectors
covered.
Visual Inspection Precision connectors should be inspected periodically. Check for the
following:
q
Bent or broken center pin
q
Damaged threads
q
Other bent or damaged connector parts
q
Dirt or foreign material in connector cavity.
C-3 REPAIR/
MAINTENANCE
ANRITSU recommends that no maintenance other than cleaning be
attempted by the customer. Any device with a suspected defective con
-
nector should be returned to ANRITSU for repair and/or service when
needed.