•Measuring Jitter Transfer Characteristics
The MP1777A can evaluate jitter transfer characteristics up to 80
MHz in applications such as 10 Gbit/s clock recovery module (O/E
converter) evaluation. Automatic (using MX177701A external soft-
ware/GPIB) and manual measurements are possible.
By combining the MP9677B E/O and O/E Converter, the MP1777A
can evaluate the jitter transfer characteristics up to 80 MHz at opti-
cal interfaces.
•Jitter Tolerance Measurement
By combining the O/E and E/O converters, the MP1777A can
measure the jitter tolerance of optical interfaces. The MX177701A
Jitter Performance Test Software and a GPIB card are required for
automatic measurement. It is also possible to perform manual
measurements without these items.
3
Data output
Clock output
DUT
GPIB
MS4630B
Network Analyzer
MP1763B
Pulse Pattern Generator
MP1777A
10 GHz Jitter Analyzer
MP9677B
E/O, O/E Converter
E/O converter
MP1763B
Pulse Pattern Generator
MP1777A
10 GHz Jitter Analyzer
MS4630B
Network Analyzer
O/E converter
Controller: Windows
®
95 or more
(with MX177701A Jitter Performance Test Software)
DUT
ATT
GPIB
MP1764A Error Detector
Controller: Windows
®
95 or more
(with MX177701A Jitter Performance Test Software)
Comentarios a estos manuales