Anritsu ME7840A Manual de usuario Pagina 46

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 126
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 45
sure that the equipment is properly connected to the bus. If
so, the following screen will be displayed:
Step 3. Click on “Run Calibration Procedure” to display a list of cali
-
bration specification files (below). (Refer to paragraph 4-3 for
a description of these files.) Running the files is described in
succeeding numbered paragraphs.
4-6 POWER LEVEL The A_Power Level.txt specification file provides prompts for a power
level calibration. Proceed as follows:
Step 1. Zero-set the power sensor for accurate measurement in fol
-
lowing steps.
Step 2. Select “Tools” then “Run Calibration Procedure.”
POWER LEVEL OPERATIONS, CALIBRATION
4-4 ME 7840A OMM
Vista de pagina 45
1 2 ... 41 42 43 44 45 46 47 48 49 50 51 ... 125 126

Comentarios a estos manuales

Sin comentarios